Anatek Labs - Moscow has recently installed the state of the art in oil and grease extraction technologies - the SPE-DEX 3000 Automated Oil and Grease Extractor from Horizon Technologies.
This new instrumentation improves accuracy, precision and sample throughput over traditional manual SPE or liquid-liquid extraction methods. For more information on this instrument, visit the Horizon Technologies Website at http://www.horizontechinc.com
Anatek Labs Inc - Moscow
Initial Performance Data and Method Detection Limit Study - EPA 1664A
| IDC |
|
|
|
MDL |
|
|
|
|
|
|
|
|
|
| Run |
HEM |
% Recovery |
|
Run |
HEM |
|
| 1 |
38.0 |
95.0 |
|
1 |
4.0 |
|
| 2 |
39.8 |
99.5 |
|
2 |
3.8 |
|
| 3 |
39.3 |
98.3 |
|
3 |
4.0 |
|
| 4 |
37.4 |
93.5 |
|
4 |
4.1 |
|
|
|
|
|
5 |
3.8 |
|
| * Spiked at 40 mg in 1L reagent water |
|
6 |
3.9 |
|
|
|
|
|
7 |
4.1 |
|
|
|
|
|
|
|
|
|
|
|
|
Std Dev = 0.13 mg/L |
|
|
|
|
|
MDL = 0.40 mg/L |
|
|
|
|
|
* Spiked at 4.0 mg in 1L reagent water |
|
|
|
|
|
|
|